SIO Argo Site:
Diagnostic/Engineering Data Page    



Serial= 3237 WMO= 5906942

Return

Link to WHOI figures

Link to Fleet Monitoring (Coriolis)

Field 1 Field 2 Field 3 Field 4

←PreviousNext→

Cycle 19 GMT 12/11/2024 21:11: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 12 12 14 0 207 14963 17 7.24 15.35 15.25 10.50 10.81 11.20


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 188 0 4.30 319 0.28 0.08 0.31 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1005.96 3.433 34.2314 1008.36 3.435 34.2310 2008.92 2.033 34.5232 0.72 16.951 33.9060


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2891 2961 0 0 0 0 1 0 2 0



Cycle 18 GMT 3/11/2024 1:48:59

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 12 11 11 0 134 14705 17 7.24 15.35 15.25 10.51 10.79 11.17


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 187 0 4.90 319 -0.24 -0.08 0.14 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1008.96 3.424 34.2312 1010.80 3.416 34.2330 2003.08 2.021 34.5236 0.60 18.038 34.1114


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2892 2947 0 0 0 0 1 0 2 0



Cycle 17 GMT 24/10/2024 6:26: 2

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 11 13 14 0 172 14537 17 7.23 15.34 15.25 10.54 10.81 11.19


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
111 165 300 184 0 0.50 318 0.12 0.04 0.32 59 1.1


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1003.72 3.432 34.2282 1005.60 3.429 34.2288 2004.52 2.028 34.5196 0.36 17.599 7.3598


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2887 2974 0 0 0 0 1 0 2 0



Cycle 16 GMT 14/10/2024 11:11: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 13 12 12 0 135 14324 17 7.24 15.36 15.26 10.47 10.78 11.16


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
111 165 300 197 0 3.00 318 -0.04 0.08 0.28 59 0.9


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1007.64 3.413 34.2286 1009.88 3.412 34.2298 2009.04 2.010 34.5218 0.56 18.223 34.0274


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2893 2951 0 0 0 0 1 0 2 0



Cycle 15 GMT 4/10/2024 15:51: 2

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 12 13 15 0 261 14583 17 7.25 15.36 15.25 10.52 10.74 11.14


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 200 0 -0.20 318 -0.08 0.12 0.28 59 0.9


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1017.48 3.381 34.2322 1019.60 3.380 34.2328 2005.52 2.021 34.5168 0.48 19.623 33.9888


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2887 2987 0 0 0 0 1 0 2 0



Cycle 14 GMT 24/ 9/2024 20:32: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 13 12 17 0 304 15185 17 7.25 15.35 15.26 10.53 10.70 11.09


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 200 0 2.10 315 -0.08 0.12 0.24 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1007.32 3.398 34.2264 1009.56 3.396 34.2270 2010.88 2.016 34.5158 0.56 21.306 34.0328


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2899 2966 0 0 0 0 1 0 2 0



Cycle 13 GMT 15/ 9/2024 0:34: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 12 13 14 0 171 14429 17 7.25 15.37 15.28 10.56 10.71 11.09


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 192 0 5.50 316 0.08 -0.04 0.30 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1001.76 3.412 34.2204 1004.84 3.403 34.2226 2001.16 2.033 34.5088 0.56 21.084 33.9082


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2882 2947 0 0 0 0 1 0 2 0



Cycle 12 GMT 5/ 9/2024 5:17: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 13 12 12 0 153 14525 17 7.24 15.37 15.27 10.57 10.71 11.09


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
111 165 300 201 0 5.00 315 -0.04 0.04 0.24 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1005.72 3.401 34.2182 1007.84 3.393 34.2206 2003.40 2.029 34.5064 0.56 21.216 33.8530


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2881 2939 0 0 0 0 1 0 2 0



Cycle 11 GMT 26/ 8/2024 9:51: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 12 13 18 0 371 14751 17 7.26 15.38 15.28 10.56 10.69 11.06


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 196 0 6.50 315 -0.04 0.04 0.24 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1003.32 3.401 34.2118 1006.28 3.396 34.2138 2001.16 2.009 34.5034 0.56 22.131 33.7140


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2891 2954 4 1 0 0 1 0 2 0



Cycle 10 GMT 16/ 8/2024 12:21: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 13 17 17 5 238 14639 17 7.25 15.38 15.27 10.60 10.69 11.06


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 187 1 -24.60 313 0.24 -0.04 0.40 59 0.9


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1002.68 3.398 34.2040 1006.16 3.388 34.2066 2003.80 2.029 34.4916 0.60 22.042 33.6738


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2887 2941 0 0 0 0 1 0 2 0



Cycle 9 GMT 6/ 8/2024 16:39: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 12 13 14 0 143 11568 17 7.26 15.40 15.30 10.59 10.74 11.13


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 185 1 -33.30 309 -0.28 0.00 0.18 59 1.0


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1010.44 3.365 34.2002 1014.00 3.357 34.2034 2005.32 2.022 34.4848 0.72 19.172 33.6318


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2868 2925 0 0 0 0 1 0 2 0



Cycle 8 GMT 27/ 7/2024 18:52: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 13 10 10 0 120 11704 17 7.26 15.40 15.30 10.62 10.72 11.09


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 185 1 -26.60 300 0.00 0.08 0.16 59 0.9


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1011.52 3.361 34.1852 1016.24 3.348 34.1918 2001.12 2.038 34.4690 0.64 20.479 33.6104


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2852 2913 0 0 0 0 1 0 2 0



Cycle 7 GMT 17/ 7/2024 21:34: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 10 8 8 0 104 10172 17 7.24 15.39 15.29 10.62 10.80 11.17


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 176 0 0.20 300 0.08 0.00 0.20 59 1.2


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1002.60 3.398 34.1596 1002.36 3.358 34.1670 1687.00 2.327 34.3926 0.60 17.599 33.5426


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2472 2553 6 1 0 0 1 0 1 0



Cycle 6 GMT 16/ 7/2024 20:51: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 8 10 10 0 155 5263 17 7.24 15.48 15.37 10.60 10.80 11.15


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
75 165 300 184 1 -10.80 301 0.08 0.12 0.16 59 1.9


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1001.56 3.374 34.1534 1004.80 3.379 34.1530 1146.08 3.068 34.2136 0.64 17.477 33.5204


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
1798 1854 0 0 0 0 1 0 1 0



Cycle 5 GMT 15/ 7/2024 21:29: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 10 10 11 1 176 9932 17 7.25 15.39 15.29 10.63 10.80 11.17


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
110 165 300 184 1 -22.60 297 -0.04 0.00 0.10 59 1.2


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
998.00 3.384 34.1342 1000.24 3.396 34.1336 1708.00 2.307 34.3694 0.48 17.553 33.5264


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2501 2593 5 1 0 0 1 0 1 0



Cycle 4 GMT 14/ 7/2024 20:47: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 9 9 9 0 128 7650 17 7.25 15.44 15.34 10.62 10.79 11.17


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
92 165 300 178 1 -18.70 294 -0.12 -0.04 0.11 59 1.6


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1000.40 3.382 34.1186 1000.84 3.377 34.1214 1408.00 2.648 34.2778 0.68 17.496 33.5114


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
2126 2179 0 0 0 0 1 0 1 0



Cycle 3 GMT 13/ 7/2024 20:45: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 9 8 10 0 121 6773 17 7.24 15.45 15.35 10.63 10.81 11.18


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
92 165 300 193 1 -32.50 290 -0.04 0.16 0.13 59 1.6


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1002.76 3.357 34.1038 1006.24 3.396 34.1002 1304.56 2.771 34.2276 0.52 17.481 33.4988


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
1993 2059 0 0 0 0 1 0 1 0



Cycle 2 GMT 12/ 7/2024 20:58: 0

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 8 7 14 0 362 5323 17 7.23 15.48 15.38 10.61 10.80 11.15


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
74 165 300 189 2 50.50 283 -0.08 0.16 0.12 59 1.9


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
1001.08 3.396 34.0688 1001.76 3.405 34.0728 1138.24 3.035 34.1456 0.48 17.505 33.4802


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
1786 1849 0 0 0 0 1 0 1 0



Cycle 1 GMT 11/ 7/2024 21:31: 1

#Eng #Pack #SentP #ConnP #ParsP SatTme SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V0 (inHG) V1 (inHG) V2 (inHG)
6 7 3 3 4 54 4502 17 7.23 15.47 15.37 10.36 10.81 11.17


PmtoSf (s) PmSf (s) FallTm (s) FallRt (mm/s) SeekTm (s) SeekP (db) PP>TIP (s/10) SPRX (db) SPRXL (db) PCorr (db) RH (%) CPUT (C)
74 165 300 182 0 2.40 290 -0.12 -0.04 -0.02 61 2.2


Pavg0 (db) Tavg0 (C) Savg0 (psu) Pavg1 (db) Tavg1 (C) Savg1 (psu) diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu)
-999.00 -999.000 -999.0000 -999.00 -999.000 -999.0000 999.60 3.386 34.0248 0.64 17.563 33.4220


HPavgI (ma) HPmaxI (ma) Vent (s/10) VentLD ExcFlg ISRID ATSBD #BinM #SubCy #IceDt
1612 1675 0 0 0 0 1 0 0 0



Cycle 0 GMT 11/ 7/2024 10:58: 1

#Eng #Pack SatTme (s) SBscan SBstat Vcpu (V) Vpmp (V) Vple (V) V1 (inHG) V2 (inHG) RH (%) CPUT (C) SPRX (db) SPRXL (db)
6 2 0 480 17 7.28 15.63 15.58 10.54 11.20 55 17.1 -0.00 -0.04


diagP0 (db) diagT0 (C) diagS0 (psu) diagP1 (db) diagT1 (C) diagS1 (psu) PmtoSf (s) PmSf (s) HPavgl (ma) HPmaxl (ma) ExcFlg ISRID AbrtCd
70.04 11.474 33.457 0.32 17.175 7.406 100 240 372 393 0 0 0


BTVcpu (V) BTvac (inHG) BTVple (V) BTavgl (mA) BTPsec (s) BTPvac (inHG) Vent (s/10) VentLD
7.28 9.78 15.57 230 405 7.56 0 0



Cycle -1 GMT 10/ 6/2024 21: 2: 1

#Eng #Pack BTPCor (db) BTVcpu (V) BTVpmp (V) BTVple (V) BTavgl (mA) BTPsec (s) BTOVcb BTOVca BTvac (inHG) BTvacI (inHG) BTRH (%) BTCPUT (C)
6 4 0.14 7.28 15.67 15.57 230.00 405 2 244 9.77 10.90 54 28.4


BTSpt (txt) BTOV BTCV BTInt Fail
"____0.05,_40.5086,__0.0000____" 0 0 6 0